Low pressure acid dissolution advantages:

Low pressure acid dissolution disadvantages:

What will and will not dissolve with this method

Most minerals <60 mesh dissolve completely, including:

A few minerals do not readily dissolve, and if they are of importance in your samples this acid digestion procedure will not be completely effective:

WARNING! At elevated temperatures the closed Teflon vessels are under pressure (a few bars)! The containers are not guaranteed by anyone under these conditions, and they may fail catastrophically! Ours have never failed catastrophically, but take heed because the possibility remains and a rupture could have awful consequences if you ignore safety precautions. Containers must be heated and cooled in a closed drying oven in a hood. Don't even peek at them unless they are at room temperature! Take adequate safety precautions, and use these methods at your own risk! These containers WILL leak at 120°C unless in a retainer clamp.

 

Note The Teflon containers used here have numbered bases and lids. Always match lid and base numbers.

Dish washing (if needed)

Clean enough 17 ml (or other size) Savillex® containers for all necessary blanks, standards, and samples, taking into account necessary replicates. With the normal dissolution procedure, each Teflon container will contain enough solution to prepare 14 final analytical solutions.

  1. Clean an appropriate number of 17 ml Teflon containers by boiling in 20% reagent grade HNO3. Cool to room temperature and rinse in DI water.
  2. Rinse in tap water (no detergent) an appropriate number of retainer clamp assemblies. Dry them in air on a paper towel.
  3. Fill the Teflon containers 60% full with a mixture of 75% water, 20% HNO3, and 5% HF. Securely tighten the covers and clamp them in retainer clamps in sets of 2 to 5. Put the clamped containers in a drying oven in the hood overnight at 120°C.
  4. In the morning, turn off the oven and let the samples cool to room temperature before opening the oven door.
  5. Remove the retainer clamp, discard the acid mixture, and thoroughly rinse the containers, including caps, with DI water.

Internal Standard Solution

To a 100 ml volumetric flask add 4 ml each of the internal standard elements Rh, In, Re, and Bi. Add 3.5 ml of high-purity HNO3 and dilute to volume. This internal standard solution is enough for up to 500 samples, and has the following concentrations:

 

HNO3 5%
Internal standards 40 ppm

Dissolution Procedure Part 1

  1. Weigh 0.1000 +/-0.0002 g of sample into a Teflon container. Wipe any dust off the rim of the vial and screw on the cap.
  2. Add 2 ml of HF and 1 ml of HNO3, and 0.1 ml of the internal standard solution. Cap the vials, clamp them into retainers, and heat in an oven at 120°C overnight (oven must be in the hood!). This step decomposes most silicates and glass and gets rid of most silica in the sample. This drastically lowers the silica activity in the second step and so aids dissolution of hard-to-dissolve silicates like tourmaline and kyanite.
  3. Cool and remove vessels from the retainers, and evaporate to dryness. We have a home-made Teflon drying oven for this purpose.

Dissolution procedure Part 2

  1. Add 3 ml of HF. Cap the vials, clamp them into retainers, and heat in an oven at 120°C overnight again (oven must be in the hood!).
  2. Cool the samples, remove vessels from the retainers, and evaporate to dryness.
  3. Add 3 ml of HNO3 and evaporate the samples to dryness. This step decomposes most insoluble fluorides and drives off excess fluoride as HF. Enough fluoride remains to keep Ti, Nb, Ta, and similar elements in solution.

Dissolution procedure Part 3

  1. Mix a solution containing high-purity acids by volume: 25% HNO3 and 5% HCl, 0.2% HF. For example, to a 500 ml bottle add 125 ml HNO3, 25 ml HCl, and 1 ml HF, and quickly fill the rest with DI water.
  2. Add 15 ml of this to each sample container.
  3. Cap the vials, clamp them into retainers, and heat in an oven at 100°C overnight (shake once after 1 hour; oven must be in the hood!). This solution has:
Dissolved solids assuming no silica lost 0.65%
Dissolved solids assuming 50% silica lost 0.33%
Dissolved solids assuming 75% silica lost 0.16%
Internal standards 261 ppb

Dilution Procedure

Take 0.5 ml of the solution from the Teflon vial and transfer it to a 13 ml plastic autosampler test tube. Add 12 ml of 1% HNO3 solution, cap, and shake. This analytical solution contains:

 

Dissolved solids assuming no silica lost 0.026%
Dissolved solids assuming 50% silica lost 0.013%
Dissolved solids assuming 75% silica lost 0.007%
HNO3 1.4%
Internal standards 10.5 ppb

 

Note that the counting times on the internals standards can be short. Note also that the concentrations of many elements in common rocks may be >100 ppb in this solution (e.g., Sr, Ba, LREE, V, Cr, Ni). This means the solution can be diluted more, or you must do the detector cross calibration, or you must analyze low abundance isotopes.