Barium and the lanthanides are relatively strong oxide formers. That is, ions like Ba+ and Ce+ react with oxygen in the plasma to form small proportions of BaO+ and CeO+ . The dominant masses of the oxide ions are 16 mass units higher than the isotope of the metal in the oxide. In the figure below, you can see that Ce has 4 naturally occurring isotopes (red bars) at masses 136, 138, 140, and 142. 140Ce is the most abundant isotope and is the one usually used for quantitative analysis. Pr has only 1 naturally occurring isotope, 141Pr (blue bars). These 5 M+ metal ions ions combine with 16O to form from 5 different mass oxide ions (MO+); for example: 136Ce+ becomes 136Ce16O+ at mass 152 (= 136 + 16).

 


Figure 1. Example metal (M+) and metal oxide (MO+) production.

 

Naturally there are also small contributions at other masses from 17O and 18O, which are 1 and 2 mass units higher than the 16O oxide. Hydroxides, such as 136Ce16OH+ at mass 153, can also be important. Neither of these is shown in Figure 1.

 

Figure 1 shows that CeO+ at mass 152 interferes with 152Sm analysis, and CeO+ at mass 158 interferes with 158Gd analysis. Oxides shown at masses 154, 156, and 157 are unimportant in this example because they are not used for quantitative analysis. There is no interference shown for 159Tb, although small quantities of 141Pr18O+, 142Ce17O+, and 142Ce16OH+ do interfere at this mass. Clearly for quantitative analysis of the metals, we must have a method of stripping away the interferences from the M+ signals.

 


Figure 2. In this example, a sample with chondritic Ba and lanthanide abundances is analyzed using pure standards. Uncorrected, the result would be the irregular red line rather than the actual black line.


Figure 3. In this more extreme example, light REE enrichment causes more severe oxide interferences on less abundant heavier REE. This, coupled with much higher Ba concentrations, leads to a substantial divergence of the uncorrected red curve compared to the correct black curve.


Figure 4. In this last example, strong light REE enrichment and very high Ba concentrations results in very severe middle REE interferences: on Sm and Eu from BaO, and from lanthanide oxides for Eu and all others. It should be pretty obvious by this time that corrections are vital for producing reasonably smooth chondrite-normalized REE patterns.

 

The simplest way to determine the size of the interferences at the masses analyzed is to measure them directly by analysis of pure, single element solutions. Then, all you have to know is where to expect the interferences, see how big they are, and calculate how to subtract them from the total signal at each mass of interest. Table 1 shows just such a set of single solution analyses.

 

Table 1. Data were obtained by analyzing 15 single element solutions (Ba-Lu), representing the data columns. The rows show signal intensities at each analyzed mass for each of the 15 solutions (signals in cps, corrected for internal standard drift, blanks subtracted). Blank subtraction yields a few negative values. The solution concentrations were 100 ppb for Ba, and 5 ppb for all other elements.
Element Analyzed mass 15 single element solutions Interfering oxides of:
Ba La Ce Pr Nd Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Ba 137 943713 116 -280 1620 459 -290 -298 -302 443 1467 559 1101 -320 -266 584  
La 139 206 287139 3 4 2 2 1 -2 31 3 3 1 1 0 0  
Ce 140 89 9 301215 4 -3 2 -1 -1 0 -1 0 0 -1 -1 -3  
Pr 141 10 0 5 341379 13 10 6 1 1 0 0 0 0 -1 -1  
Nd 146 22 0 0 0 56574 1 1 0 0 -1 0 0 0 0 -1 Ba
Eu 151 1253 1 1 1 0 0 149253 0 2 0 1 1 1 0 1 Ba
Sm 152 1576 0 8 0 -1 83404 3 44 0 -1 0 0 2 0 1 Ba, Ce
Gd 158 10 0 991 198 1437 0 0 65207 3 -2 2 0 -1 -1 -1 Ce, Pr, Nd
Tb 159 7 -1 40 12 668 1 0 0 269797 3 0 0 2 0 1 Ce, Pr, Nd
Dy 163 4 0 2 1 23 196 0 1 0 67946 1 0 0 -1 -1 Nd, Sm
Ho 165 1 0 1 0 6 184 1 -1 1 11 258167 9 7 0 2 Nd, Sm
Er 166 1 0 4 0 313 99 3 0 1 0 3 81733 2 0 1 Nd, Sm
Tm 169 2 0 1 1 2 14 100 1 2 0 9 3 243443 4 2 Sm, Eu, Gd
Yb 174 2 0 1 1 0 0 0 819 2 1 1 0 1 77287 3 Gd, Dy
Lu 175 7 -2 13 0 -1 1 -1 47 3332 -1 -1 0 -1 4 218701 Gd, Tb
Hf 178 196 0 2 -1 -1 -1 -1 1 0 460 1 3 0 0 0 Dy, Er
Ta 181 -1 0 0 -1 -1 -1 -1 0 0 8 1691 0 0 0 0 Dy, Ho
Yellow boxes are the isotopes of each element used for quantitative analyses.
Blue boxes show the signals at masses expected and observed to have oxide and hydroxide interferences ions from the element.
Interference by 159Dy16OH+ on 175Lu is expected but not observed because it is so small.

 

In Table 1, taking the ratio of the MO+ signals in the blue boxes to the M+ signals in the immediately overlying yellow boxes in Table 1 yields MO+/M+ ratios, shown in Table 2. The work needed to derive these MO+/M+ ratios is considerable, and so it is done rarely. The MO+/M+ ratios in Table 2 are referred to hereafter as the oxide calibration values.

 

Table 2. MO+/M+ ratios from the oxide calibration intensities in Table 1. For the explanation of the importance of the blue value, see text below and Table 4.
Element Mass 15 single element solutions Interfering oxides of:
Ba La Ce Pr Nd Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Ba 137                                
La 139                                
Ce 140                                
Pr 141                                
Nd 146 0.00002                             Ba
Eu 151 0.00132                             Ba
Sm 152 0.00167   0.00002                         Ba, Ce
Gd 158     0.00329 0.00058 0.02540                     Ce, Pr, Nd
Tb 159     0.00014 0.00004 0.01174                     Ce, Pr, Nd
Dy 163         0.00041 0.00235                   Nd, Sm
Ho 165         0.00010 0.00222                   Nd, Sm
Er 166         0.00560 0.00119                   Nd, Sm
Tm 169           0.00018 0.00067 0.00002               Sm, Eu, Gd
Yb 174               0.01256   0.00001           Gd, Dy
Lu 175               0.00073 0.01235             Gd, Tb
Hf 178                   0.00677   0.00004       Dy, Er
Ta 181                   0.00012 0.00655         Dy, Ho

 

Once we have the values in the table above, we can calculate the size of the oxide or hydroxide interference as follows:

 


Equation 1.

 

Ioxide,cps = Intensity of the oxide interference (in cps) during actual quantitative analysis.

 

MO+cps,cal = Oxide (or hydroxide) interference intensity (in cps) determined during oxide interference calibration (e.g., 142Ce16O+ at 158Gd+).
M+cps,cal = M+ intensity (in cps) during oxide interference calibration at the mass used for quantitative analysis (e.g., 140Ce), determined during oxide interference calibration. Note that the mass used for quantitative analysis does not have to be the mass producing the interference.
M+cps,analyzed = M+ intensity determined during actual quantitative analysis of unknown samples.

 

The [MO+cps,cal/M+cps,cal] ratios are therefore anempirical (and dimensionless) correction factors that can be used for unknown samples later on (Table 2). The absolute instrument sensitivity can vary while still preserving the proportion of oxides produced.

 

There is one problem with this method and that is that the MO+/M+ ratios are not constant; they vary with a wide variety of conditions including the purity of the argon carrier gas, the spray chamber temperature, the nebulizer gas flow, the nebulizer efficiency, and so on. Therefore, the correction factors themselves need adjustment each time lanthanides are analyzed. The graph below shows two calibration runs done to determine MO+/M+ as illustrated above. Although the two different runs had MO+/M+ ratios different by a factor of ~1.5, the changes in all of the interferences was approximately the same: ~1.5. As a result, all of the MO+/M+ ratios lie close to a straight line. Therefore, if we know the slope of this line on any given day during a run of unknowns, then we can correct the correction factors and so determine accurately the sizes of the interferences.

 


Figure 5. Original calibration (red line) vs. today's calibration (black line and blue dots). The difference between them is simply the difference in line slopes.

 

This is easily done by first analyzing a pure solution of a lanthanide, such as 146Nd+, and its oxide at 158Gd+. You may notice that 158 is not 16 mass units higher than 146.This is because the oxide at 158Gd+ is from 142Nd16O+ (mass 158), not from 146Nd16O+ (mass 162). It does not matter that our correction factor is derived from 146Nd+ rather than from the Nd isotope actually producing the oxide because the Nd isotope abundances are constant in nature (or nearly so).

 

What we need to do now is analyze a pure ~5 ppb Nd solution before running our unknowns, and measure the signal for 146Nd+ and 158Gd+ (= 142Nd16O+). This new 142Nd16O+/146Nd+ ratio will probably be different from the ratio determined during an earlier oxide calibration, as determined above.

 


Equation 2.

 

C = Correction factor, corrects all of today's MO+/M+ proportions to those measured during today's oxide calibration, according to the linear relationship shown in the figure above.
Numerator = Today's determined NdO+/Nd+ ratio, determined at masses 158 and 146, respectively.
Denominator = NdO+/Nd+ ratio, determined originally during the oxide calibration procedure. These values are also measured at masses 158 and 146, respectively.

 

Now we have the MO+/M+ proportions for all oxide and hydroxide interferences of Ba, the lanthanides, Hf, and Ta (determined some time ago during the oxide calibration), and we have the correction factor (determined today). During the analysis of unknowns, the signals on the analyzed element masses (yellow in Table 1) are used to calculate the oxide interference on all pertinent higher masses (blue in Table 1) by the equation:

 


Equation 3.

 

These values can then be subtracted from the signals at the masses having the interferences to yield the actual signal produced by M+ ions, stripped of the oxide interferences:

 


Equation 4.

 

Icorrected,cps = Intensity of the M+ signal at a given mass, in cps, after removal of the interferences.
Ioriginal,cps = Intensity of the total signal at a given mass, in cps, before removal of the interferences..

 

The ellipsis at the end of Equation 4 indicates additional interference correction like that in the curved brackets but for additional oxide species. For example, Nd and Sm produce oxides that interfere with 163Dy. Referring to Tables 1 and 2:

  1. The interference from Nd on mass 163 is 23/56574 = 0.00041, based on the Table 1 oxide calibration.
  2. The interference from Sm on mass 163 is 196/83404 = 0.00235, based on the Table 1 oxide calibration.

Using correction factor "C", which is determined each day using the pure Nd solution is constant for any given analysis procedure on a given day. The actual intensity of the mass 163 peak, due exclusively to 163Dy is:

 

163Dy = (original mass 163 intensity) - (C * 0.00041 * intensity of 142Nd) -(C * 0.00235 * intensity of 152Sm)  Equation 5.

 

These concepts are all well and good, but what do you need to really do before analyzing a bunch of samples?

 

Table 3. Calculating the C factor.
Item Example value Notes
Today's measured Nd/Gd (A) 0.01875 Measured today on a ~5 ppb Nd solution
Original calibration Nd/Gd (B) 0.02540 The calibration in bold blue elsewhere on this page
Today's correction factor: A/B = C  0.738 The C value that you type in to the interference corrections in the software, shown in Table 4

 

Table 4 shows all of the oxide correction equations for Ba, the lanthanides, Hf, and Ta. The equations are more-or-less in the format that they appear in the Elan 6100 software, correction equations section. The coefficients vary over time as updates are made.

 

Table 4. Interference correction equations including isotope overlaps (not discussed on this page) and the oxide interference corrections. In red is the C factor, which is also the numerator of Equation 2, and in blue is the denominator of equation 2. The part C factor should be determined prior to each analytical run for lanthanides, Hf, and Ta, and typed directly into the Elan corrections table. Note that equation 5, above, is identical to the Dy row correction.
Element Mass Isotope overlap corrections Oxide interference corrections
Ba 137    
La 139    
Ce 140    
Pr 141    
Nd 146   - ((C * 0.00002) * 137Ba)
Eu 151   - ((C * 0.00132) * 137 Ba)
Sm 152 - (0.00805 * 158Gd) - ((C * 0.00167) * 137 Ba) - ((C * 0.00002) * 140Ce)
Gd 158 - (0.00402 * 163Dy) - ((C * 0.00329) * 140Ce) - ((C * 0.00058) * 141Pr) - ((C * 0.02540) * 146Nd)
Tb 159   - ((C * 0.00014) * 140Ce) - ((C * 0.00004) * 141Pr) - ((C * 0.01174) * 146Nd)
Dy 163   - ((C * 0.00041) * 146Nd) - ((C * 0.00235) * 152Sm)
Ho 165   - ((C * 0.00010) * 146Nd) - ((C * 0.00222) * 152Sm)
Er 166   - ((C * 0.00560) * 146Nd) - ((C * 0.00119) * 152Sm)
Tm 169   - ((C * 0.00018) * 152Sm) - ((C * 0.00067) * 151Eu) - ((C * 0.00002) * 158Gd)
Yb 174   - ((C * 0.01256) * 158Gd) - ((C * 0.00001) * 163Dy)
Lu 175 - (0.00594 * 178Hf) - ((C * 0.00073) * 158Gd) - ((C * 0.01235) * 159Tb)
Hf 178   - ((C * 0.00677) * 163Dy) - ((C * 0.00004) * 166Er)
Ta 181   - ((C * 0.00012) * 163Dy) - ((C * 0.00655) * 165Ho)

 

Although this oxide correction procedure is an approximation, it has worked quite well in practice. Note that there is a simpler way to enter the C correction in the Elan software, but in practice it does not work reliably.